HES Series - Powerful​

  • Input voltage up to 1,200 V
  • Load current up to 1350 A (higher currents available upon request)
  • Power output from 600 W to 14,400 W (higher power outputs available upon request)
  • Temporary power overload capacity, depending on model
  • Models with integrated zero-volt supply (ZV models)
  • Models with several current setting and measuring ranges (MR models)
  • High power density up to 2,100 W in 2 U (3,200 W in 3 U)
  • Basic operating modes CC – CV – CR – CP
  • Combined operating modes CC+CV, CR+CC+CV, CP+CC+CV, CV+CC
  • Ethernet + USB + RS-232 + CAN as standard
  • GPIB optional
  • Intuitive operation via touchscreen
  • MPP tracking
    Energy storage testing
    Internal resistance measurement
  • Modulator function
    Square wave/PWM function
  • Dynamic load profiles with synchronous data acquisition
  • Trigger model
  • SCPI programming with measurement function
  • Measurement data storage directly to a USB flash drive
  • Adjustable thresholds for overcurrent and undervoltage
  • Analog measurement outputs for voltage and current
  • Analog control input
  • Digital input and programmable control output
  • Digital master-slave operation for performance expansion
  • Fan cooling with stepless fan control, power- and current-dependent
  • Save settings
    Electronic protection
    Bilingual help system (DE, EN)

Description

H&H’s All-Rounder: the HES Series.
A wide range of models combines high power density with extensive standard features. The devices can be operated locally via the user-friendly interface or remotely via the integrated interfaces. Programming is performed via SCPI with an extensive command set.

Zero-Volt Models – PLIxxxxZV

Special applications, such as fuel cell tests, need to load the DUT down to almost 0 V. For these purposes we supply a range of models with an integrated zero-volt supply. When I/O port signals are to be used, the GND of the I/O port must be galvanically isolated from the negative load input. Therefore we recommend to install the optional galvanically isolated I/O port (option PLI06) at these models.

Multi-Range Models – PLIxxxxMR

A model variant with 2, 3 or 4 “real” current ranges allows the setting and measurement of current, power and resistance over a wide range of values with high resolution and accuracy. Especially solar and PV applications have high requirements at this point. The ranges are graduated with factor 10 each.

Operating Modes and Functions

Basic and Combined Operating Modes

The devices have the operating modes constant current (CC), constant voltage (CV), constant resistance (CR) and constant power (CP). Furthermore, protection values for undervoltage and overcurrent can be set in each operating mode. This results in the combined operating modes CC+CV, CR+CC+CV, CP+CC+CV, CV+CC.

Dynamic Lastprofile with List  Function

Dynamic operation (LIST function) can be simulated with up to 300 setpoints. The electronic load stores the synchronously recorded measurement values along with a timestamp. An individual sampling rate can be defined for each setpoint. This allows critical time intervals to be sampled at a higher resolution.

Data Acquisition

An integrated data acquisition function allows to store measurement data internal or directly on an external USB flash drive.

Energy Storage Test/Discharge Function

The discharge function discharges energy storage devices such as accumulators, batteries, ultracaps etc. in a controlled manner under definable switch-off conditions. The values determined in the process, such as charge and energy, can be read locally and via a data interface.

Internal Resistance Measurement

The electronic load can measure the DC internal resistance of the connected test object. For this purpose it uses a calculation method which is specified in various standards, e.g. DIN EN 61951, DIN EN 61960.

Modulator Function

The modulator function enables the superimposed modulation of a setpoint (e.g., current or voltage) with an AC signal. This allows dynamic superimpositions to be simulated. Parameters such as frequency, amplitude, and waveform are configured.

Square Wave/PWM Function

The square wave or PWM function can be used to generate periodic load changes with definable frequency, duty cycle, and amplitude. This function is particularly suitable for simulating switched loads and for investigating the dynamic behavior of power supplies.

Master-Slave Operation

To increase the power or current, up to 5 equal devices can be operated in parallel in a master-slave connection. The system behaves in the network externally like one single device. The master unit controls the total current of the system, displays the total measured values and supplies these when queried via one of the data interfaces. Cabling: One set each of K-MS-PLI and K-MS-CAN master-slave cables on all slave units (to be purchased from H&H or assembled by the user). To be able to tap monitor signals etc. when using the master-slave cable K-MS-PLI, we offer a SubD25 coupler as an accessory. (Limited scope of functions in master-slave mode. Control via CAN interface not possible).

Watchdog Function

If the watchdog function is activated, the load input is switched off in case of faulty communication with the control PC in order to protect the connected DUT.

MPP Tracking (option PLI21)

The Maximum Power Point Tracking (MPPT) function is used to test solar or photovoltaic modules. For this purpose the maximum power point (MPP) of the connected DUT is controlled in CV mode.

Trigger Model

A configurable trigger model controls data acquisition, static and dynamic settings.

Loading Capacity

The model range covers power classes from  600 W, up to 28,000 W continuous power rating. In addition the models up to 300 V have an overload capability (See Technical Specifications). The height and duration of the possible overload depends on the temperature of the power stage. The unit indicates the currently possible overload. Therefore the units can be used  even for considerably more powerful applications.

Protections

  • Current limiting
  • Undervoltage limiting
  • Power limiting
  • Overtemperature protection
  • Protection of GND lines on the analog I/O port

Data Interfaces

The following data interfaces are included as standard:

  • Ethernet
  • USB
  • RS-232
  • CAN

Optionally deliverable:

  • GPIB (option PLI02)

LabVIEW drivers are certified by National Instruments and are available for download on the NI website (IDNet).

Analog I/O Port

The standard analog I/O port provides the following functions:

  • Load settings C and V (CR and CP indirectly via µC)
  • Analog limit settings (overcurrent, undervoltage)
  • Load input control
  • Operating mode selection
  • Monitor signals for load current and voltage
  • Trigger input
  • Trigger output
  • Digital input
  • Programmable digital output
  • Status outputs
  • Control speed selection
  • Remote shunt down

Galvanically Isolated Analog I/O Port

For the galvanic isolation between the Analog I/O Port and the load terminals the PLI06 option can be installed. Isolation voltage is 800 V. Using this board also prevents ground loops.

Help System

A help system for most menu items is integrated in the user interface. The help system’s language can be set to German or English.

Calibration 

A free Factory Calibration Certificate (FCC) is supplied with the devices. The calibration process is subject to supervision in accordance with DIN EN ISO 9001. The calibration certificate documents the traceability to national standards to illustrate the physical device in accordance with the international System of Units (SI). For use under laboratory conditions, H&H recommends a calibration interval of 2 years. This is an empirical value that can be used as a guide for the first period of use. Depending on the intended use, service life, relevance of the application and ambient conditions, the operator should adjust this interval accordingly.

Documentation

We supply a user manual as pdf file and printed General Safety Instructions, each in German and English.

PLI Standard Model overview

ProduktUmaxImaxPmax
HES1120660 V1050 A11200 W
HES11212120 V525 A11200 W
HES11230300 V240 A11200 W
HES11260600 V160 A11200 W
HES11280800 V120 A11200 W
HES12630300 V270 A12600 W
HES12660600 V180 A12600 W
HES12680800 V135 A12600 W
HES1280660 V1200 A12800 W
HES12812120 V600 A12800 W
HES14030300 V300 A14000 W
HES140660 V120 A1400 W
HES14060600 V200 A14000 W
HES1406C2060 V20 A1200 W
HES14080800 V150 A14000 W
HES1412120 V40 A1400 W
HES1430300 V32 A1400 W
HES1440660 V1350 A14400 W
HES14412120 V675 A14400 W
HES1460600 V16 A1400 W
HES1480800 V12 A1400 W
HES1600660 V1500 A16000 W
HES16012120 V750 A16000 W
HES210660 V180 A2100 W
HES210660 V180 A2100 W
HES2106C3060 V30 A1800 W
HES2112120 V60 A2100 W
HES2130300 V48 A2100 W
HES2160600 V24 A2100 W
HES2180800 V18 A2100 W
HES320660 V300 A3200 W
HES3212120 V150 A3200 W
HES3230300 V60 A3200 W
HES3260600 V40 A3200 W
HES3280800 V30 A3200 W
HES4230300 V90 A4200 W
HES4260600 V60 A4200 W
HES4280800 V45 A4200 W
HES480660 V450 A4800 W
HES4812120 V225 A4800 W
HES5630300 V120 A5600 W
HES5660600 V80 A5600 W
HES5680800 V60 A5600 W
HES60660 V60 A600 W
HES606C1060 V10 A600 W
HES612120 V20 A600 W
HES630300 V16 A600 W
HES640660 V600 A6400 W
HES6412120 V300 A6400 W
HES660600 V8 A600 W
HES680800 V6 A600 W
HES7030300 V150 A7000 W
HES7060600 V100 A7000 W
HES7080800 V75 A7000 W
HES800660 V750 A8000 W
HES8012120 V375 A8000 W
HES8430300 V180 A8400 W
HES8460600 V120 A8400 W
HES8480800 V90 A8400 W
HES960660 V900 A9600 W
HES9612120 V450 A9600 W
HES9830300 V210 A9800 W
HES9860600 V140 A9800 W
HES9880800 V105 A9800 W
ProduktUmaxImaxPmax
HES24K12HV1200 V12 A2400 W